Table is Sticking due to Cradle Release Block is Extended.Table suddenly sticks and/or stops when moving out of the bore…

Table is Sticking due to Cradle Release Block is Extended.Table suddenly sticks and/or stops when moving out of the bore.Perform Cradle Release Block Retraction Adjustment-wqr.Discovery MR750 20.1 IB2

Symptom

Table suddenly sticks and/or stops when moving out of the bore. Table suddenly sticks and/or stops when moving out of the bore Table suddenly sticks and/or stops when moving out of the bore.

Error message

2268911 Cradle clutch slip detected out of tolerance, motion was stopped Error: 2268911 Host: lx-mr Process: NSP File: SCP:LongMotionMonit Line: 1805 Exception Class: Unknown Severity: Unknown Function: Patient Handling: Control/Display PSD: FSE-XL Coil: HNS Head Scan: 2649/9 Cradle clutch slip detected out of tolerance, motion was stopped Expected Velocity: 100000 um/s Position Velocity: 4976 um/s Motor Speed: 25216 um/s-

2268909 The system has detected more than expected clutch slip during cradle motion, motion may be impaired Tue Nov 18 15:42:11 2008 2268909 1SLMAINMR NSPSCP:LongMotionMonit 1842Exception Class: Unknown Severity: UnknownFunction: Patient Handling: Control/DisplayPSD: Spin Echo Coil: HEAD Scan: 50113/1The system has detected more than expected clutch slip during cradle motion, motion may be impaired.Expected Velocity: 100000 um/sPosition Velocity: 13970 um/sMotor Speed: 45216 um/s->

Applies to

Models
Discovery
Product
Discovery MR750. Also called DVMR. This product is compatible with 3.0T 20.0, 20.1_IB1, IB2, 21.0 and DV22.0 software only
Software
20.1_IB1 and IB2 This release of software is applicable to Discovery MR450 and Discovery MR750 systems. See Long Description 20.1_IB1 is the unified Discovery MR450 and Discovery MR750 software release. Known internally to engineering as "DVIB". This release of software is applicable to Discovery MR450 and Discovery MR750 systems. 20.1_IB2 is the upgrade (bug fixes) for IB1.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.