Cannot scan with Error 2267863 The % axis XGA is reporting a I_Coil Wire Fault.

Symptom

Cannot scan with Error: 2267863 The % axis XGA is reporting a I_Coil Wire Fault Cannot scan with Error: 2267863 The % axis XGA is reporting a I_Coil Wire Fault.

Error message

2267863 The %axis XGA is reporting a I_Coil Wire Fault See Long Description Extended Error: The cables between the coil current sensor and the SB board is loose or has fallen off. Ensure cables J5 and J6 of the XGA are connected to the corresponding axis current sensor at the top of the Power Cabinet.

Applies to

Models
Discovery
Product
Discovery MR450 Also called DVMR. This product is compatible with 1.5T 20.1, 20.1_IB1, IB2, 21.0 and DV22.0 software systems only.
Software
20.1_IB1 and IB2 This release of software is applicable to Discovery MR450 and Discovery MR750 systems. See Long Description 20.1_IB1 is the unified Discovery MR450 and Discovery MR750 software release. Known internally to engineering as "DVIB". This release of software is applicable to Discovery MR450 and Discovery MR750 systems. 20.1_IB2 is the upgrade (bug fixes) for IB1.
Component
5184566-2 XGD AMPLIFIER 5184566-2 XGD AMPLIFIER
Area
MR

Full cause & step-by-step resolution