GE Healthcare · MROptima22675172267446

Unable to Scan with Message The IRF3 is not ready due to Fiber Optic Transceiver in %-axis Port of XGA Control Bd in Cro…

Unable to Scan with Message The IRF3 is not ready due to Fiber Optic Transceiver in %-axis Port of XGA Control Bd in Cronus chassis.Replace 5183329-2 Fiber Optic Transceiver SFP 4Gb Multimode-wqr.Opti…

Symptom

Unable to Scan with Message The IRF3 is not ready. See Long Description Customer is unable to scan due to message in message window located at bottom of Scan Desktop which states “The IRF3 is not ready.->”

Error message

2267517 Calibrations were aborted because the IRF3 driver found problems during system initialization. 2267517 Host: nbomr2 Process: diagCoord File: irdCal.cpp Line: 122 Task ID: 0x493d650 Time: 9096 Object: AGP Exception Class: Notification Function: Software/Firmware/Config: N/A Calibrations were aborted because the IRF3 driver found problems during system initialization. Calibration will not run if there are communication or configuration problems with the RRX, DTX, VRF, or VRE.->

2267446 The XGD module did not send a Fiber Link Init Complete

Applies to

Models
Optima
Product
Optima MR450W Wide bore magnet. Not Discovery750 or Discovery450.
Software
21.0 Only for Optima MR450W (wide bore) systems. (Not to be used on Discovery MR450 or Discovery MR750). Only for Optima MR450W (wide bore) systems. (Not to be used on Discovery MR450 or Discovery MR750).
Component
5183329-2 Fiber Optic Transceiver SFP 4Gb Multimode, RoHS-Compliant Note, as of March 24, 2010 this part can not be ordered yet. It will be a FRU. If you cant order the transceiver order the higher FRU.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.