GE Healthcare · MRDiscovery226891110000025216226883615402011531771

2268911 Cradle clutch slip detected due to improper dock centering.Advance to scan is Required Multiple times to get Tab…

2268911 Cradle clutch slip detected due to improper dock centering.Advance to scan is Required Multiple times to get Table to Scan Position.Perform Dock Centering in Light Weight Patient Transport Adj…

Symptom

Advance to scan is Required Multiple times to get Table to Scan Position

Error message

2268911 Cradle clutch slip detected out of tolerance, motion was stopped Error: 2268911 Host: lx-mr Process: NSP File: SCP:LongMotionMonit Line: 1805 Exception Class: Unknown Severity: Unknown Function: Patient Handling: Control/Display PSD: FSE-XL Coil: HNS Head Scan: 2649/9 Cradle clutch slip detected out of tolerance, motion was stopped Expected Velocity: 100000 um/s Position Velocity: 4976 um/s Motor Speed: 25216 um/s-

2268836 Cradle move failed to reach target position within position tolerance. Error: 2268836 Host: lx-mr Process: NSP File: SCP:LongMotionMonit Line: 2095 Exception Class: Unknown Severity: Unknown Function: Patient Handling: Control/Display PSD: FSE-XL Coil: HNS Head Scan: 2649/9 Cradle move failed to reach target position within position tolerance Target: 1540201 um Acutal: 1531771 um Tolerance: 500 um

Applies to

Models
Discovery
Product
Discovery MR750. Also called DVMR. This product is compatible with 3.0T 20.0, 20.1_IB1, IB2, 21.0 and DV22.0 software only
Software
20.0 Also called DVMR. This software is compatible with 3.0T Signa MR750 systems only
Area
MR