GE Healthcare · MRDiscovery
RF hub will not power on due to No Shunt connector is on J3 of the power distribution board.
Symptom
RF hub will not power on RF hub will not power on.
Applies to
- Models
- Discovery
- Product
- Discovery MR450 Also called DVMR. This product is compatible with 1.5T 20.1, 20.1_IB1, IB2, 21.0 and DV22.0 software systems only.
- Software
- 20.1_IB1 and IB2 This release of software is applicable to Discovery MR450 and Discovery MR750 systems. See Long Description 20.1_IB1 is the unified Discovery MR450 and Discovery MR750 software release. Known internally to engineering as "DVIB". This release of software is applicable to Discovery MR450 and Discovery MR750 systems. 20.1_IB2 is the upgrade (bug fixes) for IB1.
- Component
- 6250054 POWER DISTRIBUTION BOARD 6250054 POWER DISTRIBUTION BOARD. Not interchangeable with 5250082 board.
- Area
- MR
Full cause & step-by-step resolution