GE Healthcare · CTDiscovery26013470779382279400558796050189418827
The table fails to move up or down but the elevation readout numbers on the gantry display do change whenever a move button is pressed and somethimes the readout shows a negative value.
Symptom
The table fails to move up or down but the elevation readout numbers on the gantry display do change whenever a move button is pressed and somethimes the readout shows a negative value.
Error message
Error: 260134707 Elevation could not stop with stop command. If table is moving downward, check valve See Long Description for error versions Mon Sep 19 08:02:03 2005 Error: 260134707 Host: tgp Process: tgp_subsys File: gtcb_emc_movechk.cxx Line: 616 Mon Sep 19 08:02:03 2005 Host : Table Sub-System Ermes # : 260134707 Exception Class : Secondary Severity : Pri/Soft File : gtcb_emc_movechk.cxx Line# : 616 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 Elevation could not stop with stop command. If table is moving downward, check valve. last position = 793822 um current position = 794005 um commanded direction = UP time from stop = 587960 ms Mon Dec 12 10:22:24 2005 260134707 Host: tgp tgp_subsys gtcb_emc_movechk.cxx 579 Mon Dec 12 10:22:24 2005 Host : Table Sub-System Ermes # : 260134707 Exception Class : Secondary Severity : Pri/Soft File : gtcb_emc_movechk.cxx Line# : 579 Function : No System Function Reported Scan Type : None/Unknown Scan: 50189/1/0 Elevation could not stop with stop command. If table is moving downward, check valve. last position = 418827 um current position = 418382 um commanded direction = DOWN time from stop = 1040 ms Error: 2601
Applies to
- Models
- Discovery
- Product
- Discovery CT750 HD V40_PS_HD64_G_GTL V40_PS_HD64_G_GTM Discovery HD CT750 Method of visual identification HD gantry: Gray top and Gray Side covers with HD Markings Console Type: GOC6 has black SCIM and tower face Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of softwares the product can run (Consult attached FCT Software Cross Reference): 08MW39.4 - M3 (see attachement for M3 available options) 09MW29.14 - M4 (HD Options = M3 plus Gemstone Spectral Imaging, Gemstone Spectral Review and volume Helical Shuttle ) Example: Of difference with options software [insite@ct01 ~]$ swokinstall -p Product Software Option Keys: Gemstone Spectral Imaging Permanent HelicalShuttle Permanent Xtream Recon Permanent FAQ's Table weight limit: 500 lbs Tube Type - D3702T Green D3701T Current production - M4 Refer to product data sheets or Technical reference Manual, English 5317222-1EN Revision : 2 or later Refer to Application Tips and Work-Arounds Direction 5311815-1EN Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 5127774 ELEVATION POTENTIOMETER ASSY-VCT See Long Description for details Part #: 5127774 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: ELEVATION POTENTIOMETER ASSY FRU: Yes Status: Orderable Comments: source pole/planner code revised per to Global Table transfer