Heat Exchanger Cabinet (HEC) detected the customer coolant flow is reversed Due to Vendor Chiller Water Flow Issue.

Symptom

Coolant flow Error Does Not Stop Scan 2261707 Heat Exchanger Cabinet (HEC) detected the customer coolant flow is reversed Coolant flow Error Does Not Stop Scan 2261707 Heat Exchanger Cabinet (HEC) detected the customer coolant flow is reversed.

Error message

2261707 Heat Exchanger Cabinet (HEC) detected the customer coolant flow is reversed. This alarm will not stop scanning see long Error: 2261707 Host: THDMRI03 Process: NSP File: SCP:CoolingHWO Line: 2957 Exception Class: Unknown Severity: Unknown Function: Software/Firmware/Config: N/A PSD: FRFSE-XL Coil: HDCTL 234 Scan: 57/3 Heat Exchanger Cabinet (HEC) detected the customer coolant flow is reversed. Please ensure the facility supply and return coolant hoses are correctly hooked up to the HEC This alarm will not stop scanning

Applies to

Models
Optima
Product
Optima MR450W Wide bore magnet. Not Discovery750 or Discovery450.
Software
21.0 Only for Optima MR450W (wide bore) systems. (Not to be used on Discovery MR450 or Discovery MR750). Only for Optima MR450W (wide bore) systems. (Not to be used on Discovery MR450 or Discovery MR750).
Area
MR