Pular para o conteúdo principal
GE Healthcare · MROptima2267827

XGD Hardware ovrSR Errors Logged but System Functioning Well due to In Some Scans Gradient + ECC Compensation will Excee…

XGD Hardware ovrSR Errors Logged but System Functioning Well due to In Some Scans Gradient + ECC Compensation will Exceed Slew-Rate Limit and DSP Firmware will Detect and Report a Warning.

Symptom

XGD Hardware ovrSR Errors Logged but System Functioning Well See Long Description We see these errors being logged during scanning and also during Gradient Hammer diagnostic but there is no apparent effect on system operation. The system seems to be running well. What do these errors mean and should we troubleshoot the system because of them? Sat Feb 14 07:43:22 2009 Error: 2267827 Host: SLMAINMR Process: NSP File: SCP:FspMac.cpp Line: 320 Exception Class: Unknown Severity: Unknown Function: Gradient: N/A PSD: Gradient Echo Coil: 8HRBRAIN Scan: 543/2 XGD hardware warning detected. The ovrSR bit in DSP_FAULT register was detected on the XGD XGA-CB board X axis. Sat Feb 14 07:43:22 2009 Error: 2267827 Host: SLMAINMR Process: NSP File: SCP:FspMac.cpp Line: 320 Exception Class: Unknown Severity: Unknown Function: Gradient: N/A PSD: Gradient Echo Coil: 8HRBRAIN Scan: 543/2 XGD hardware warning detected. The ovrSR bit in DSP_FAULT register was detected on the XGD XGA-CB board Z axis. These errors sometimes seen when running Gradient Hammer diagnostic: Tue Feb 24 22:21:21 2009 Error: 2267827 Host: SLMAINMR Process: diagCoord File: FspMac.cpp Line: 320 Diagnostic: Gradient Hammer Iteration: 1 of 1 Action: 1 Task ID: 0x2549508 Time: 548284 Object: SCP Function: Gradient: N/A XGD hardware warning detected. The ovrSR bit in DSP_FAULT register was detected on the XGD XGA-CB board X axis. Tue Feb 24 22:21:21 2009 Error: 2267827 Host: SLMAINMR Process: diagCoord File: FspMac.cpp Line: 320 Diagnostic: Gradient Hammer Iteration: 1 of 1 Action: 1 Task ID: 0x2549508 Time: 548334 Object: SCP Function: Gradient: N/A XGD hardware warning detected. The ovrSR bit in DSP_FAULT register was detected on the XGD XGA-CB board Z axis.

Error message

2267827 XGD hardware warning detected. The %s bit in %s register was detected on the %b %b. See Long Description Error: 2267827 Host: SLMAINMR Process: NSPFile: SCP:FspMac.cpp Line: 320Exception Class: Unknown Severity: UnknownFunction: Gradient: N/APSD: Gradient Echo Coil: 8HRBRAIN Scan: 543/2 XGD hardware warning detected. The ovrSR bit in DSP_FAULT register was detected on the XGD XGA-CB board % axis.

Applies to

Models
Optima
Product
Optima MR450W Wide bore magnet. Not Discovery750 or Discovery450.
Software
21.0 Only for Optima MR450W (wide bore) systems. (Not to be used on Discovery MR450 or Discovery MR750). Only for Optima MR450W (wide bore) systems. (Not to be used on Discovery MR450 or Discovery MR750).
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.