GE Healthcare · MROptima227355122735502275612
Fails TPS Reset when booting up to Applications or when Resetting TPS Due to the RF Control Board.
Symptom
Fails TPS Reset when booting up to Applications or when Resetting TPS See Long Description When the MR system is booting up to applications it will fail TPS reset but Applications will successfully boot. With Applications up and "Reset TPS" has been selected, TPS fails to download successfully.
Error message
2273551 RF Hub Power Failure. The RF Hub detected a power failure on the following board. RF Hub (board name), power +-%V. 2273550 RF Hub Power Failure see long Error: 2273550 Host: WBH_RAD_MR11 Process: NSP File: SCP:RfHubHw::Pdb Line: 734 Exception Class: Unknown Severity: Unknown Function: RF: RF Receive PSD: Unknown Coil: Unknown Scan: -1/-1 RF Hub Power Failure. The RF Hub RFCB read that the PDB is reporting power failures for the following power distributions. Common: OK RFCB: -10V MCDB: -10V RFSB: OK-> 2275612 RF Hub Hardware Failure. 2275612 NSP SCP:RfHubCanHWO 4310 Exception Class: Unknown Severity: Unknown Function: RF: RF Receive PSD: Gradient Echo Coil: HNS_ CTLTOP Scan: 3/1 RF Hub Hardware Failure. A problem with the RF Hub hardware was detected. TPS reset and scanning will fail until the problem is fixed. Check for previous errors in the error log for more information.->
Applies to
- Models
- Optima
- Product
- Optima MR450W Wide bore magnet. Not Discovery750 or Discovery450.
- Software
- 21.0 Only for Optima MR450W (wide bore) systems. (Not to be used on Discovery MR450 or Discovery MR750). Only for Optima MR450W (wide bore) systems. (Not to be used on Discovery MR450 or Discovery MR750).
- Component
- 5250098 1.5T RF Hub Control Board
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA ID100513. Will Not Scan with Gradient Errors.
- CWA ID103384. Computer Will Not Boot with the message on the screen that says Do Not Interact With System.
- CWA ID103582. Unable To Scan With Driver Module Errors.
- CWA ID104077. Computer Will Not Boot with the message on the screen that says Do Not Interact With System.
- IApplication Software does not complete the boot process with TPS Download failure Due to Idle Overnight With Power On.
- CWA ID107471. Unable to Download or Reset The TPS MGD or ISE.