Pular para o conteúdo principal
GE Healthcare · MRDiscovery22740212274013

Multicoil Bias Fault errors on multiple coils due to loose screws inside front of Hypertonics coil connector.

Symptom

Multicoil Bias Fault errors on multiple coils.

Error message

2274021 Coil Fault - Short Circuit A multicoil bias fault was detected See Long Description Error: 2274021 Host: WBH_RAD_MR11 Process: NSP File: SCP:RfHubCanHWO::RfBias Line: 4080 Exception Class: Unknown Severity: Unknown Function: RF: RF Bias PSD: Spin Echo Coil: HNS CTL123 Scan: 4637/1 Coil Fault - Short Circuit A multicoil bias fault was detected. Connector: Port 2 (P2) Channels faulted: 0x100 First channel: 9 of 32, counting from 1 Fault value: -4777 mV, Channel: 9

2274013 Function RF Bias Unused Channel Fault - Short Circuit A multicoil bias fault was detected see long 2274013 Function RF Bias Unused Channel Fault - Short Circuit A multicoil bias fault was detected Connector: Port A Channels faulted: 0x10 First channel: 5 of 8, counting from 1 Fault value: 1551 mV, Channel: 5-> Extended Error: A short circuit was detected by the RF Hub on a channel that is not known to be connected a coil. This could indicate a coil is plugged in that is not being read by coil ID. Run the Multicoil Bias Diagnostic or Coil Datapath Diagnostic to troubleshoot. Explanation of fields: Connector - The coil connector the fault was detected on. Channels faulted - This is a hexadecimal bit map of al

Applies to

Models
Discovery
Product
Discovery MR750. Also called DVMR. This product is compatible with 3.0T 20.0, 20.1_IB1, IB2, 21.0 and DV22.0 software only
Software
20.1_IB1 and IB2 This release of software is applicable to Discovery MR450 and Discovery MR750 systems. See Long Description 20.1_IB1 is the unified Discovery MR450 and Discovery MR750 software release. Known internally to engineering as "DVIB". This release of software is applicable to Discovery MR450 and Discovery MR750 systems. 20.1_IB2 is the upgrade (bug fixes) for IB1.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues