GE Healthcare · MROptima226751222675212267520

TPS Reset failed with errors 2267512, 2267520 and 2267521 due to defective RF Switch board RFSB.

Symptom

Unable to Download or Reset the TPS MGD or ISE Unable to download TPS or ISE or MGD chassis

Error message

2267512 R1 gain calibration failed because the signal was received at the RRX for one or more channels was too small Error: 2267512 - The R1 gain calibration failed because the signal was received at the RRX for one or more channels was too small. This will cause TPS_reset to fail.->

2267521 The R1 gain calibration failed Error 2267521 - The R1 gain calibration failed because one or more channels are out of spec. This error will cause a TPS reset failure. See the file /usr/g/service/log/r1GainCal_<date>.dat.->

2267520 The R1 gain calibration failed because the signal was received at the RRX Error: 2267520 Host: LIBR-MR1 Process: LxMxConsumer File: RrfCalibrator.cpp Line: 1130 Server Name: ADS The R1 gain calibration failed because the signal was received at the RRX for one or more channels was too small. This will cause TPS_reset to fail.->

Applies to

Models
Optima
Product
Optima MR450W Wide bore magnet. Not Discovery750 or Discovery450.
Software
21.0 Only for Optima MR450W (wide bore) systems. (Not to be used on Discovery MR450 or Discovery MR750). Only for Optima MR450W (wide bore) systems. (Not to be used on Discovery MR450 or Discovery MR750).
Component
6250052 RF Hub Switch Board RFSB 1 board required for 16 channel system, 2 boards required for 32 channel systems.
Area
MR