Error Code 2261714 is Being Logged Periodically due to the HEC Signal Box. 2261714 is Being Logged Periodically.
Symptom
2261714 is Being Logged Periodically. This is Heat Exchanger Cabinet (HEC) detected the gradient coil coolant resistivity has reached a warning level. See Long Description The error message shown below is being logged periodically. It's being logged every few minutes. What does this mean and how do we get it to stop being logged? Mon Sep 15 05:55:48 2008 2261714 LX-BAY11B NSP SCP:CoolingHWO 2957 Exception Class: Unknown Severity: Unknown Function: Software/Firmware/Config: N/A PSD: Unknown Coil: Unknown Scan: -1/-1 Heat Exchanger Cabinet (HEC) detected the gradient coil coolant resistivity has reached a warning level. Please contact service. This alarm will not stop scanning->
Error message
2261714 Heat Exchanger Cabinet (HEC) detected the gradient coil coolant resistivity has reached a warning level. Please contact service. This alarm will not stop scanning See Long Description Mon Sep 15 05:55:48 2008 2261714 LX-BAY11B NSP SCP:CoolingHWO 2957 Exception Class: Unknown Severity: Unknown Function: Software/Firmware/Config: N/A PSD: Unknown Coil: Unknown Scan: -1/-1 Heat Exchanger Cabinet (HEC) detected the gradient coil coolant resistivity has reached a warning level. Please contact service. This alarm will not stop scanning-> Extended Error: Order coolant deionization kit and schedule service. If the resistivity is 0 kOhm-cm, check the wiring between the sensor in the gradient coolant reservoir and the signal box.
Applies to
- Models
- Discovery
- Product
- Discovery MR750. Also called DVMR. This product is compatible with 3.0T 20.0, 20.1_IB1, IB2, 21.0 and DV22.0 software only
- Software
- 20.1_IB1 and IB2 This release of software is applicable to Discovery MR450 and Discovery MR750 systems. See Long Description 20.1_IB1 is the unified Discovery MR450 and Discovery MR750 software release. Known internally to engineering as "DVIB". This release of software is applicable to Discovery MR450 and Discovery MR750 systems. 20.1_IB2 is the upgrade (bug fixes) for IB1.
- Component
- 5269691 Signal Box kit Signal Box kit 5269691
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA ID100513. Will Not Scan with Gradient Errors.
- Corrupt Image Database.
- CWA ID103384. Computer Will Not Boot with the message on the screen that says Do Not Interact With System.
- CWA ID103582. Unable To Scan With Driver Module Errors.
- CWA ID104077. Computer Will Not Boot with the message on the screen that says Do Not Interact With System.
- CWA ID104279. System Software suddenly locked up.