Pular para o conteúdo principal
GE Healthcare · MRDiscovery2267836

2267836 The Z axis XGA is reporting a cold plate over-temperature The cold-plate temperature is 59C due to defective XGA…

2267836 The Z axis XGA is reporting a cold plate over-temperature The cold-plate temperature is 59C due to defective XGA 5184566-2.Discovery MR750 20.1_IB2

Symptom

Fails TPS Reset when booting up to Applications or when Resetting TPS See Long Description When the MR system is booting up to applications it will fail TPS reset but Applications will successfully boot. With Applications up and "Reset TPS" has been selected, TPS fails to download successfully.

Error message

2267836 The Z axis XGA is reporting a cold plate over-temperature The cold-plate temperature is 59C. see long Error: 2267836 Host: DVMR Process: NSP File: SCP:FspMac.cpp Line: 320 Exception Class: Unknown Severity: Unknown Function: Gradient: N/A PSD: Unknown Coil: Unknown Scan: -1/-1 The Z axis XGA is reporting a cold plate over-temperature. The cold-plate temperature is 59C.->

Applies to

Models
Discovery
Product
Discovery MR750. Also called DVMR. This product is compatible with 3.0T 20.0, 20.1_IB1, IB2, 21.0 and DV22.0 software only
Software
20.1_IB1 and IB2 This release of software is applicable to Discovery MR450 and Discovery MR750 systems. See Long Description 20.1_IB1 is the unified Discovery MR450 and Discovery MR750 software release. Known internally to engineering as "DVIB". This release of software is applicable to Discovery MR450 and Discovery MR750 systems. 20.1_IB2 is the upgrade (bug fixes) for IB1.
Component
5184566-2 XGD AMPLIFIER 5184566-2 XGD AMPLIFIER
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.