Loud noise as system prepares to make xray, or start or brake the rotor.

Symptom

Loud noise as system prepares to make xray, or start or brake the rotor. Loud noise as system prepares to make xray for a tube warmup, scout or helical scan

Error message

Error: 260118177 Jedi (40-0114)Rotor not at speed. See Long Description for complete error Thu Feb 15 14:07:54 2007 260118177 4 orp orp_subsys gen_exec.c 2376 Thu Feb 15 14:07:54 2007 Host : Jedi Ermes # : 260118177 Exception Class : Abort Severity : Primary File : gen_exec.c Line# : 2376 Function : No System Function Reported Scan Type : Static Scan: 50054/1/1 (40-0114) Rotor not at speed . Z6,Z7= 2h,0h Mon Apr 17 14:01:27 2006 Error: 260118177 Host: orp Process: orp_subsys File: gen_exec.c Line: 2150 Mon Apr 17 14:01:26 2006 Host : Jedi Ermes # : 260118177 Exception Class : Abort Severity : Primary File : gen_exec.c Line# : 2150 Function : No System Function Reported Scan Type : Static Scan: 50148/1/1 (40-0114)Rotor not at speed. ^C Z6,Z7= 2h,0h

Applies to

Models
Discovery
Product
Discovery CT750 HD V40_PS_HD64_G_GTL V40_PS_HD64_G_GTM Discovery HD CT750 Method of visual identification HD gantry: Gray top and Gray Side covers with HD Markings Console Type: GOC6 has black SCIM and tower face Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of softwares the product can run (Consult attached FCT Software Cross Reference): 08MW39.4 - M3 (see attachement for M3 available options) 09MW29.14 - M4 (HD Options = M3 plus Gemstone Spectral Imaging, Gemstone Spectral Review and volume Helical Shuttle ) Example: Of difference with options software [insite@ct01 ~]$ swokinstall -p Product Software Option Keys: Gemstone Spectral Imaging Permanent HelicalShuttle Permanent Xtream Recon Permanent FAQ's Table weight limit: 500 lbs Tube Type - D3702T Green D3701T Current production - M4 Refer to product data sheets or Technical reference Manual, English 5317222-1EN Revision : 2 or later Refer to Application Tips and Work-Arounds Direction 5311815-1EN Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
D3701T : PERSEUS TUBE UNIT - Used on CT750 HD See Long Description for details. Part #: D3701T Part Type: GP TUBES Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: PERSEUS TUBE UNIT (GREEN) FRU: Yes Status: Orderable Comments: Per HDVCT RSPL, 212008898, 8/19/08. ** FE ASSIST ** Important to have adequate HV oil in the HV Wells. Making sure some oil does overflow when inserting Candle stick. (212001261 4/6/09)
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required