Start Exam button not available after selecting protocol Duet to Table Can't Find Home Position.

Symptom

Start Exam button not available after selecting protocol See Long Description for details Start Exam button not available after selecting protocol. Site selects and accepts the protocol, but still can't get Start Exam button at lower right of screen to activate.

Error message

2268831 Move request not allowed when table is not docked and fully up See Long Description 2268831 NSP SCP:SriManager 2991 Exception Class: Unknown Severity: Unknown Function: Patient Handling: Control/Display PSD: Unknown Coil: Unknown Scan: -1/-1 Move request not allowed when table is not docked and fully up

Applies to

Models
Discovery
Product
Discovery MR450 Also called DVMR. This product is compatible with 1.5T 20.1, 20.1_IB1, IB2, 21.0 and DV22.0 software systems only.
Software
20.1_IB1 and IB2 This release of software is applicable to Discovery MR450 and Discovery MR750 systems. See Long Description 20.1_IB1 is the unified Discovery MR450 and Discovery MR750 software release. Known internally to engineering as "DVIB". This release of software is applicable to Discovery MR450 and Discovery MR750 systems. 20.1_IB2 is the upgrade (bug fixes) for IB1.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.