Pular para o conteúdo principal
GE Healthcare · MRDiscovery2268905226890450171

Unable to Scan with RF Air Temperature Faults Due To Customer Coolant Chiller.

Symptom

Unable to Scan with RF Air Temperature Faults

Error message

2268905 RF air temperature fault detected. Scanning will be inhibited until the RF air temperature matches expected. See Long Description 2268905 NSP SCP:SriCanHWO 1998 Exception Class: Unknown Severity: Unknown Function: Patient Handling: Control/Display PSD: Unknown Coil: Unknown Scan: -1/-1 RF air temperature fault detected. Scanning will be inhibited until the RF air temperature matches expected Temperature difference: 107 0.1 degrees C Tolerance: 30 0.1 degrees C-> Extended Error: The Heat Exchanger (HEC) manages RF air temperature. This message indicates that the HEC was unable to achieve the requested RF air temperature.

2268904 RF air temperature warning detected. Future scanning will be inhibited until RF air temperature matches expected. See Long Description Thu Jan 15 14:56:24 2009 Error: 2268904 Host: signa2 Process: NSP File: SCP:SriCanHWO Line: 2030 Exception Class: Unknown Severity: Unknown Function: Patient Handling: Control/Display PSD: Gradient Echo Coil: HEAD Scan: 50171/1 RF air temperature warning detected. Future scanning will be inhibited until RF air temperature matches expected. Temperature difference: 20 0.1 degrees C Tolerance: 20 0.1 degrees C-> Extende

Applies to

Models
Discovery
Product
Discovery MR450 Also called DVMR. This product is compatible with 1.5T 20.1, 20.1_IB1, IB2, 21.0 and DV22.0 software systems only.
Software
20.1_IB1 and IB2 This release of software is applicable to Discovery MR450 and Discovery MR750 systems. See Long Description 20.1_IB1 is the unified Discovery MR450 and Discovery MR750 software release. Known internally to engineering as "DVIB". This release of software is applicable to Discovery MR450 and Discovery MR750 systems. 20.1_IB2 is the upgrade (bug fixes) for IB1.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues