GE Healthcare · MRDiscovery2254058
Defective XRFD RF Amplifier causing error 2254058.
Symptom
Will Not Scan With RF Errors Will not scan with RF errors
Error message
Error: 2254058 RF Amplifier in low power state; avoid MEMP, FSE scans. A RF Amplifier has successfully bypassed a fault. IQ may be compromised. Error: 2254058 Host: MNMR750 Process: NSP File: SCP:Srfd3RfNbCanHWO Line: 2296 Exception Class: Unknown Severity: Unknown Function: RF: RF Transmit PSD: Unknown Coil: Unknown Scan: -1/-1 RF Amplifier in low power state; avoid MEMP, FSE scans.^A RF Amplifier has successfully bypassed a fault. IQ may be compromised. Please run prescan again. Scanning can then continue at reduced output levels. Scanning may be limited. Contact FE.-> Scanning with MEMP, FSE PSD's may be limited, especially with large patients.
Applies to
- Models
- Discovery
- Product
- Discovery MR750. Also called DVMR. This product is compatible with 3.0T 20.0, 20.1_IB1, IB2, 21.0 and DV22.0 software only
- Software
- 20.1_IB1 and IB2 This release of software is applicable to Discovery MR450 and Discovery MR750 systems. See Long Description 20.1_IB1 is the unified Discovery MR450 and Discovery MR750 software release. Known internally to engineering as "DVIB". This release of software is applicable to Discovery MR450 and Discovery MR750 systems. 20.1_IB2 is the upgrade (bug fixes) for IB1.
- Component
- 5135844-3 3T XRFD RF Amp 5135844-3 3T XRFD RF Amp
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA ID100513. Will Not Scan with Gradient Errors.
- Corrupt Image Database.
- CWA ID103384. Computer Will Not Boot with the message on the screen that says Do Not Interact With System.
- CWA ID103582. Unable To Scan With Driver Module Errors.
- CWA ID104077. Computer Will Not Boot with the message on the screen that says Do Not Interact With System.
- CWA ID104279. System Software suddenly locked up.