Gradient error 2267894 due to failed XGD Power Supply 5148804-2, also known as XPS

Symptom

Will Not Scan with Gradient Errors.

Error message

2267894 The X axis XPS-SB is reporting an input bus over-current see long description Tue May 12 16:35:45 2009 Error: 2267894 Host: WBH_RAD_MR11 Process: NSP File: SCP:FspMac.cpp Line: 320 Exception Class: Unknown Severity: Unknown Function: Gradient: N/A PSD: Unknown Coil: Unknown Scan: -1/-1 The X axis XPS-SB is reporting an input bus over-current.-> Extended Error: Check that 420V input voltage measures correctly. If the fault persists after TPS Resets, and the 420V input power measures correctly, the XPS should be replaced

Applies to

Models
Discovery
Product
Discovery MR750. Also called DVMR. This product is compatible with 3.0T 20.0, 20.1_IB1, IB2, 21.0 and DV22.0 software only
Software
20.1_IB1 and IB2 This release of software is applicable to Discovery MR450 and Discovery MR750 systems. See Long Description 20.1_IB1 is the unified Discovery MR450 and Discovery MR750 software release. Known internally to engineering as "DVIB". This release of software is applicable to Discovery MR450 and Discovery MR750 systems. 20.1_IB2 is the upgrade (bug fixes) for IB1.
Component
5148804-2 XGD Power Supply XPS 5148804-2 XGD Power Supply XPS ** 2nd Opinion Part ** Confirm appropriate extended error messages have been read and gradient diagnostics have been executed before placing order for this part. 212002950 May 28, 2009. Promoted this part for DVMR 3.0T NPI program, 212035293, 10 April 2008.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.