Pular para o conteúdo principal
GE Healthcare · CT230009013

Lightspeed 1.X - Octane

Symptom

Multiple recon retries and suspended images during an exam. See long description for details Multiple recon retries and suspended images during an exam. To get the images to recon you would have to resume the queue or reboot the system for the images to recon.

Error message

Error: 230009013 Queue entry suspended. See Long Description for sample error. Fri May 27 08:25:44 2005 Error: 230009013 Host: ct2 Process: queuemgr File: qmgr_list_mgmt.c Line: 2275 Function: Image Generation : SBC Processing Job Type: Retro Helical Scan: 5431/2/1 Image: 5431/102/391 Queue entry suspended.

Applies to

Product
Lightspeed 1.X - Octane 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete Lightspeed 1.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config showprods|grep Apps note: if check_config does not work then the software is not up to date with all FMI's. Check FMI history and verify that all patches have been loaded since last software reload. Ref FMI 25269 List of Supported Console Types Octane with Pegasus Recon hardware. may have an 02 SBC - Run df -k to see if the sbc is partitioned. List of softwares the product can run 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3176T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2225004-9 73GB scan disk (replaces 2225004-8 and 18 GB p/n 2225004-4 and 36 GB scan disk 2225004-5) See Long Description for Details. Part #: 2225004-9 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: 73GB 15000 RPM Scan Data Drive FRU: Yes Status: Orderable Comments: Per ECR 2046173, 212008898, 8/8/07. ** FE ASSIST ** Refer to Service Note SN07110612a For Jumper information (12/3/07 212001261) Part #: 2225004-8 Part Type: GP CONSUMABLES Part Weight: 1.61 LBR Repairable: No Returnable: No Hazardous Material: N Description: HARD DRIVE 73GB 15KRPM SCAN DATA FRU: Yes Status: Orderable Comments: Added per ECR 2031409, 7/14/06, ms11892. ** FE ASSIST ** If no stock of the 2225004-8 order 2225004-9 as the replacement. Refer to Service Note SN07110612a (212001261 12/3/07)
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.