Pular para o conteúdo principal
GE Healthcare · MRDiscovery226751722238172267410129162256925

Fails TPS Reset when booting up to Applications or when Resetting TPS due to Defective SFP (fiber module).

Symptom

Fails TPS Reset when booting up to Applications or when Resetting TPS See Long Description When the MR system is booting up to applications it will fail TPS reset but Applications will successfully boot. With Applications up and "Reset TPS" has been selected, TPS fails to download successfully.

Error message

2267517 Calibrations were aborted because the IRF3 driver found problems during system initialization. 2267517 Host: nbomr2 Process: diagCoord File: irdCal.cpp Line: 122 Task ID: 0x493d650 Time: 9096 Object: AGP Exception Class: Notification Function: Software/Firmware/Config: N/A Calibrations were aborted because the IRF3 driver found problems during system initialization. Calibration will not run if there are communication or configuration problems with the RRX, DTX, VRF, or VRE.->

2223817 TPS is Not Responding. Please Reset TPS Wed Apr 28 16:58:01 2004 Error: 2223817 Host: GEMS Process: NSP File: TardisReset.cc Line: 398 TPS is not responding. Please reset TPS.

2267410 NO DTX1 MODULE FOUND Error: 2267410Host: GEMRISC1 Process: diagCoordFile: Irf3.cpp Line: 2257Task ID: 0x493ef40 Time: 12916 Object: AGPException Class: NotificationFunction: Software/Firmware/Config: N/ANO DTX1 MODULE FOUND.The IRF3 board does not see a DTX1 module connected to by dual fiber cable.The system can not function without a DTX1.

2256925 StartupCondition error on Irf3 resulted in TPS reset failure. 2256925 Host: nbomr2 Process: LxMxConsumer File: services/initialization/ServerInitManager Line: 767Ser

Applies to

Models
Discovery
Product
Discovery MR450 Also called DVMR. This product is compatible with 1.5T 20.1, 20.1_IB1, IB2, 21.0 and DV22.0 software systems only.
Software
20.1_IB1 and IB2 This release of software is applicable to Discovery MR450 and Discovery MR750 systems. See Long Description 20.1_IB1 is the unified Discovery MR450 and Discovery MR750 software release. Known internally to engineering as "DVIB". This release of software is applicable to Discovery MR450 and Discovery MR750 systems. 20.1_IB2 is the upgrade (bug fixes) for IB1.
Component
5183329-2 Fiber Optic Transceiver SFP 4Gb Multimode, RoHS-Compliant Note, as of March 24, 2010 this part can not be ordered yet. It will be a FRU. If you cant order the transceiver order the higher FRU.
Area
MR

Full cause & step-by-step resolution