GE Healthcare · MRDiscovery2261703

Error 2261703 customer coolant too high, always over 7 bar even with no flow due to defective sensor.

Symptom

2261703 Heat Exchanger Cabinet (HEC) detected customer coolant pressure is too high, always over 7 bar.

Error message

2261703 Heat Exchanger Cabinet (HEC) detected customer coolant pressure is too high. See Long Description. 2261703 NSP SCP:CoolingHWO 2836 Exception Class: Unknown Severity: Unknown Function: Software/Firmware/Config: N/A PSD: Unknown Coil: Unknown Scan: -1/-1 Heat Exchanger Cabinet (HEC) detected customer coolant pressure is too high. Please ensure the facility is providing the correct coolant flow to the HEC cabinet This alarm will not stop scanning-> Extended Error: Heat Exchanger Cabinet (HEC) detected customer coolant pressure is too high. Please ensure the facility is providing the correct coolant flow to the HEC cabinet %b

Applies to

Models
Discovery
Product
Discovery MR450 Also called DVMR. This product is compatible with 1.5T 20.1, 20.1_IB1, IB2, 21.0 and DV22.0 software systems only.
Software
20.1_IB1 and IB2 This release of software is applicable to Discovery MR450 and Discovery MR750 systems. See Long Description 20.1_IB1 is the unified Discovery MR450 and Discovery MR750 software release. Known internally to engineering as "DVIB". This release of software is applicable to Discovery MR450 and Discovery MR750 systems. 20.1_IB2 is the upgrade (bug fixes) for IB1.
Component
5267643 Liquid Pressure Sensor Kit 5267643 Liquid Pressure Sensor Kit
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.