GE Healthcare · MRDiscoveryOptimaBrivo226741222720592267597
Fails TPS Reset when booting up to Applications or when Resetting TPS Due to Defective RRX Module.
Symptom
Fails TPS Reset when booting up to Applications or when Resetting TPS See Long Description When the MR system is booting up to applications it will fail TPS reset but Applications will successfully boot. With Applications up and "Reset TPS" has been selected, TPS fails to download successfully.
Error message
2267412 Not enough RRX modules found. The system did not find enough RRX modules for the number of channels configured. Extended Error: The system needs 1 RRX per 16 channels. Each RRX is enumerated through its connection to the RF Hub. A bad cable between a RRX and the RF-hub, or a cable connected to the wrong jack of the RF Hub can cause the RRX to appear to be configured for the wrong set of channels and thus not counted in the required number of RRX. If the RF Hub is not powered up, the RRX connected to it will all default to channels 113 - 128. 2272059 VRF in ICN1 is connected to RRX with ID 3 NOTE: ID3 is invalid. System presently only recognizes ID 1 and ID 2. 2267597 The RRX % module was not detected by IRF3. The system can not scan without this module See Long Description Mon Jun 13 10:58:22 2011 Error: 2267597 Host: OWS Process: diagCoord File: Irf3.cpp Line: 2593 Task ID: 0x488b060 Time: 9441 Object: AGP Exception Class: Notification Function: Software/Firmware/Config: N/A The RRX_2 module was not detected by IRF3. The system can not scan without this module.-> Extended Error : The system needs 1 RRX per 16 channels. Each RRX is enumerated through its connection to the
Applies to
- Models
- Discovery, Optima, Brivo
- Product
- Discovery MR750. Also called DVMR. This product is compatible with 3.0T 20.0, 20.1_IB1, IB2, 21.0 and DV22.0 software only
- Software
- 20.1_IB1 and IB2 This release of software is applicable to Discovery MR450 and Discovery MR750 systems. See Long Description 20.1_IB1 is the unified Discovery MR450 and Discovery MR750 software release. Known internally to engineering as "DVIB". This release of software is applicable to Discovery MR450 and Discovery MR750 systems. 20.1_IB2 is the upgrade (bug fixes) for IB1.
- Component
- 5174169 HDv Remote Receiver Module 5174169 HDv Remote Receiver Module
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA ID100513. Will Not Scan with Gradient Errors.
- Corrupt Image Database.
- CWA ID103384. Computer Will Not Boot with the message on the screen that says Do Not Interact With System.
- CWA ID103582. Unable To Scan With Driver Module Errors.
- CWA ID104077. Computer Will Not Boot with the message on the screen that says Do Not Interact With System.
- CWA ID104279. System Software suddenly locked up.