Pular para o conteúdo principal
GE Healthcare · MRDiscoveryOptimaBrivo2267601226750033836

Unable to Download or Reset The TPS MGD or ISE Due to Defective HDv Remote Receiver Module.

Symptom

Unable to Download or Reset the TPS MGD or ISE Unable to download TPS or ISE or MGD chassis

Error message

2267601 AGP forced TPS_reset to fail because AGP could not successfully calibrate the Exciters, Receivers, and Rf-Hub. Look for calibration related error messages in this log See Long Description 2267601 ERROR: EM_IRF3_CAL_FAILED AGP forced TPS_reset to fail because AGP could not successfully calibrate the Exciters, Receivers, and Rf-Hub. Look for calibration related error messages in this log. Extended Error : TPS_reset will fail if the AGP can not calibrate the TG tables for the exciters, the DC offset of the receivers, and the R1 gain of the RF Hub.

2267500 RRX Zero DC Offset Failure 2267500 Host: THDMRI03 Process: diagCoord File: rrxCal.cpp Line: 247 Task ID: 0x47d7cf0 Time: 33836 Object: AGP Exception Class: Notification Function: Software/Firmware/Config: N/A RRX Zero DC Offset Failure One or more channels on the RRX module(s) sees a DC offset that is too large to compensate for. The file /usr/g/service/log/rrxZeroOffset<date>.log has details on which RRX modules have the DC offset overrange.->

Applies to

Models
Discovery, Optima, Brivo
Product
Discovery MR750. Also called DVMR. This product is compatible with 3.0T 20.0, 20.1_IB1, IB2, 21.0 and DV22.0 software only
Software
20.1_IB1 and IB2 This release of software is applicable to Discovery MR450 and Discovery MR750 systems. See Long Description 20.1_IB1 is the unified Discovery MR450 and Discovery MR750 software release. Known internally to engineering as "DVIB". This release of software is applicable to Discovery MR450 and Discovery MR750 systems. 20.1_IB2 is the upgrade (bug fixes) for IB1.
Component
5174169 HDv Remote Receiver Module 5174169 HDv Remote Receiver Module
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required