Pular para o conteúdo principal
GE Healthcare · MRDiscoveryOptima607871451941

Known Problem in 20.1_IB1 Software, Resolved in 20.1_IB2 or 20.1_IB3 Software.

Symptom

Lost Graphic RX Toolbar During a Series Prescription. see long description Lost Graphic RX toolbar 1. The user may experience the loss of the Graphic Rx Toolbar during a series prescription. 2. The Graphic Rx toolbar may disappear from the screen after a real-time scan (Only Lx real-time scans. For eg. SmartPrep and not MR-Echo). Workaround/Recovery: 1. End exam and start a new exam 2. Select the Graphic Rx Toolbar icon to re-open. See page 3 of service note in Resolution. Fix: 20.1_IB2 FMI 60787 PR# 1451941

Applies to

Models
Discovery, Optima
Product
Discovery MR750. Also called DVMR. This product is compatible with 3.0T 20.0, 20.1_IB1, IB2, 21.0 and DV22.0 software only
Software
20.1_IB1 and IB2 This release of software is applicable to Discovery MR450 and Discovery MR750 systems. See Long Description 20.1_IB1 is the unified Discovery MR450 and Discovery MR750 software release. Known internally to engineering as "DVIB". This release of software is applicable to Discovery MR450 and Discovery MR750 systems. 20.1_IB2 is the upgrade (bug fixes) for IB1.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues