GE Healthcare · MRDiscoveryOptima6079060787

Known Problem in 20.1_IB1 Software, Resolved in 20.1_IB2 or 20.1_IB3 Software.

Symptom

Unable to save series with PURE or ASSET with 12CH Body Array Upper coil configuration. see long description 12ch Body Array MR450 only . The Body Upper mode of the coil is not compatible with the PURE/ASSET Calibration scan, due to the inability to reduce the calibration FOV. A window appears asking the operator to set FOV to 50 cm, but after correcting they cannot save the series. An error message is seen as follows: "Prescribed series cannot be saved". The correct FOV value should be 48 cm. However, the MR scanner does not allow the user to change the value. Workaround: Request optional FMI 60790 "12 Channel Body Coil Config Update" from FMI Admin and alert OLC. An engineer from the MR Coils Integration team will then deploy the FMI remotely. See Page 3 of service note found in Resolution. Fix: 20.1_IB2 FMI 60787

Applies to

Models
Discovery, Optima
Product
Discovery MR450 Also called DVMR. This product is compatible with 1.5T 20.1, 20.1_IB1, IB2, 21.0 and DV22.0 software systems only.
Software
20.1_IB1 and IB2 This release of software is applicable to Discovery MR450 and Discovery MR750 systems. See Long Description 20.1_IB1 is the unified Discovery MR450 and Discovery MR750 software release. Known internally to engineering as "DVIB". This release of software is applicable to Discovery MR450 and Discovery MR750 systems. 20.1_IB2 is the upgrade (bug fixes) for IB1.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.