GE Healthcare · MRDiscovery2252306
Error 2252306 due to shorted RF transmit cable.
Symptom
Unable To Scan With Driver Module Errors Unable To Scan With Driver Module Errors.
Error message
2252306 TR Driver Fault Detected Short Circuit Source of Fault is Head TR See Long Description for Extended Error. 2252306 ERROR: EM_DCB_TR_SHORT_CKT TR Driver Fault Detected - Short Circuit. Source of Fault is %b. Value: %d mA Limit: %d mA Extended Error: If this error happens repeatedly then check for a short-circuited PIN diode or other short-circuit condition somewhere in the failing transmit chain. Use TR DD MC System Path diagnostic or Gather Load Data and verify that the short circuit values are less than 30mA. If error happens intermittently and cannot be traced to failed hardware then it may be due to gradient noise in the system.
Applies to
- Models
- Discovery
- Product
- Discovery MR750. Also called DVMR. This product is compatible with 3.0T 20.0, 20.1_IB1, IB2, 21.0 and DV22.0 software only
- Software
- 20.0 Also called DVMR. This software is compatible with 3.0T Signa MR750 systems only
- Component
- 5184360 Cable track assembly 5184360 Cable track assembly-Functions. The A Port and other associated cables take-up is located on the right side of the rear pedestal. Its FRU is noted as Cable Track Assembly-Functions.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA ID100513. Will Not Scan with Gradient Errors.
- Corrupt Image Database.
- CWA ID103384. Computer Will Not Boot with the message on the screen that says Do Not Interact With System.
- CWA ID103582. Unable To Scan With Driver Module Errors.
- CWA ID104077. Computer Will Not Boot with the message on the screen that says Do Not Interact With System.
- CWA ID104279. System Software suddenly locked up.