The Scan Stops Because the Application Did Not Receive a Trigger on Time due to Defective PAC.ECG Cardiac Gating and PPG…

The Scan Stops Because the Application Did Not Receive a Trigger on Time due to Defective PAC.ECG Cardiac Gating and PPG Pulse Gating lose their Waveform as Soon as Scan Starts

Symptom

ECG Cardiac Gating and PPG Pulse Gating lose their Waveform as Soon as Scan Starts Customer also had a third party Invivo monitor for cardiac monitoring, the waveform on the Invivo monitor did not drop out during the scan

Error message

2218537 Scan Stopped due to error : Have not detected any new Cardiac Triggers for quite a while. 2218537 Host: signa2 Process: scn Scan Stopped due to error : Have not detected any new Cardiac Triggers for quite a while. Check the Cardiac waveform display to see if the Cardiac waveform is present.

Applies to

Models
Discovery
Product
Discovery MR750. Also called DVMR. This product is compatible with 3.0T 20.0, 20.1_IB1, IB2, 21.0 and DV22.0 software only
Software
20.1_IB1 and IB2 This release of software is applicable to Discovery MR450 and Discovery MR750 systems. See Long Description 20.1_IB1 is the unified Discovery MR450 and Discovery MR750 software release. Known internally to engineering as "DVIB". This release of software is applicable to Discovery MR450 and Discovery MR750 systems. 20.1_IB2 is the upgrade (bug fixes) for IB1.
Component
5112668-2 PAC–2B MODULE PAC–2B MODULE 5112668-2 PAC–2B MODULE 5112668-2