Persisted Worklist Query Failure for DR-F system with V5 and V6 software.

Symptom

Persisted Worklist Query Failure for DR-F system with V5 and V6 software. For the persisted patients, DR-F software will compare their encoding ((0008,0005) SpecificCharacterSet) with current system encoding configuration (ISO_IR 100/ISO_IR192/GB18030). So the persisted patients will not be displayed if any of below conditions is met: 1) RIS doesn’t send Tag (0008,0005) 2) RIS sends Tag (0008,0005) with NULL value 3) RIS sends Tag (0008,0005) with different value with current system encoding configuration

Applies to

Models
Brivo
Product
Brivo DR-F LING LONG DR-F
Area
X-ray R&F

Full cause & step-by-step resolution