Pular para o conteúdo principal
GE Healthcare · MRDiscovery225727150290

2257271 The Universal Power Monitor 1 detected too much power on the Hybrid Splitter (Body Channel) due to UPM calibrati…

2257271 The Universal Power Monitor 1 detected too much power on the Hybrid Splitter (Body Channel) due to UPM calibrations not performed for hybrid reflected power.

Symptom

Scan Stops with error 2257271 The Universal Power Monitor 1 detected too much power on the Hybrid Splitter (Body Channel) Scan Stops with error 2257271 The Universal Power Monitor 1 detected too much power on the Hybrid Splitter (Body Channel).

Error message

2257271 The Universal Power Monitor 1 detected too much power on the Hybrid Splitter (Body Channel) see long Error: 2257271 Host: RCMR1 Process: NSP File: SCP:UPM Line: 481 Exception Class: Unknown Severity: Unknown Function: RF: RF Power Monitor PSD: Spin Echo Coil: BODY Scan: 50290/1 The Universal Power Monitor 1 detected too much power on the Hybrid Splitter (Body Channel). Forward Power = 79. Reflected Power = 154.->

Applies to

Models
Discovery
Product
Discovery MR750. Also called DVMR. This product is compatible with 3.0T 20.0, 20.1_IB1, IB2, 21.0 and DV22.0 software only
Software
20.1_IB1 and IB2 This release of software is applicable to Discovery MR450 and Discovery MR750 systems. See Long Description 20.1_IB1 is the unified Discovery MR450 and Discovery MR750 software release. Known internally to engineering as "DVIB". This release of software is applicable to Discovery MR450 and Discovery MR750 systems. 20.1_IB2 is the upgrade (bug fixes) for IB1.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues