Pular para o conteúdo principal
GE Healthcare · MRDiscovery22676113577157

The RF Hub reports that there is no LO signal on the cable from the exciter/clock ref module to the RF Hub control board…

The RF Hub reports that there is no LO signal on the cable from the exciter/clock ref module to the RF Hub control board Due to the 3.0T RF Hub Control Board.

Symptom

Unable to Download or Reset the TPS MGD or ISE Unable to download TPS or ISE or MGD chassis

Error message

2267611 The RF Hub reports that there is no LO signal on the cable from the exciter/clock ref module to the RF Hub control board See Long Description Error - Tue Jul 14 02:32:39 2009 Error: 2267611 Host: C-3TMRI2 Process: diagCoord File: rfHub.cpp Line: 638 Diagnostic: RF Calib Iteration: 2 of infinite Action: 1 Task ID: 0x7f38530 Time: 3577157 Object: AGP Test Task Function: Software/Firmware/Config: N/A The RF Hub reports that there is no LO signal on the cable from the exciter/clock ref module to the RF Hub control board.-> Possible cause(s): NB DTx Module, RF Hub Control Board

Applies to

Models
Discovery
Product
Discovery MR750. Also called DVMR. This product is compatible with 3.0T 20.0, 20.1_IB1, IB2, 21.0 and DV22.0 software only
Software
20.1_IB1 and IB2 This release of software is applicable to Discovery MR450 and Discovery MR750 systems. See Long Description 20.1_IB1 is the unified Discovery MR450 and Discovery MR750 software release. Known internally to engineering as "DVIB". This release of software is applicable to Discovery MR450 and Discovery MR750 systems. 20.1_IB2 is the upgrade (bug fixes) for IB1.
Component
5250098-2 3.0T RF HUB CONTROL BOARD 5250098-2 3.0T RF HUB CONTROL BOARD. Comments: MUST run all receiver diagnostics, including Coil Datapath Diagnostic, prior to placing order for this part. 212002950 October 24, 2008. Part is being set up for 3.0T DVMR program RSPL. 501139061, 5-Mar-08.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.