Attention The detector is not at the proper temperature.
Symptom
Attention The detector is not at the proper temperature. see long description for more detail. Attention The detector is not at the proper temperature. If you wish to proceed with this scan select the "OK" button, however IQ may be degraded including image artifacts. Please contact GE Service
Error message
Error: 260132913 Plenum Plenum Filtered Average Temperature Out of Tolerance See long description for full error Error: 260132913 Plenum Plenum Filtered Average Temperature Out of Tolerance Wed Jan 18 19:50:57 2006 Error: 260132913 Host: orp Process: orp_subsys File: ThermalManager.cxx Line: 2381 Wed Jan 18 19:50:50 2006 Host : Das Ermes # : 260132913 Exception Class : Secondary Severity : Pri/Soft File : ThermalManager.cxx Line# : 2381 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 Plenum Plenum Filtered Average Temperature Out of Tolerance Sensor Temperatures: Left: 2850, Center: 2750, Right: 2850 Filtered Average Temperature: 2538 Limits: 2600, 3400 Thu Jan 19 22:49:08 2006 Error: 260132913 Host: orp Process: orp_subsys File: ThermalManager.cxx Line: 2381 Thu Jan 19 22:49:06 2006 Host : Das Ermes # : 260132913 Exception Class : Secondary Severity : Pri/Soft File : ThermalManager.cxx Line# : 2381 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 Plenum Plenum Filtered Average Temperature Out of Tolerance Sensor Temperatures: Left: 3000, Center: 2850, Right: 2750 Filtered Average Temperature: 2555 Limits: 2600, 3400 This error
Applies to
- Models
- Discovery
- Product
- Discovery CT750 HD V40_PS_HD64_G_GTL V40_PS_HD64_G_GTM Discovery HD CT750 Method of visual identification HD gantry: Gray top and Gray Side covers with HD Markings Console Type: GOC6 has black SCIM and tower face Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of softwares the product can run (Consult attached FCT Software Cross Reference): 08MW39.4 - M3 (see attachement for M3 available options) 09MW29.14 - M4 (HD Options = M3 plus Gemstone Spectral Imaging, Gemstone Spectral Review and volume Helical Shuttle ) Example: Of difference with options software [insite@ct01 ~]$ swokinstall -p Product Software Option Keys: Gemstone Spectral Imaging Permanent HelicalShuttle Permanent Xtream Recon Permanent FAQ's Table weight limit: 500 lbs Tube Type - D3702T Green D3701T Current production - M4 Refer to product data sheets or Technical reference Manual, English 5317222-1EN Revision : 2 or later Refer to Application Tips and Work-Arounds Direction 5311815-1EN Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Area
- CT
Full cause & step-by-step resolution