GE Healthcare · CT189057189060
CT/e, CT Prospeed AI/FI Irix or Linux
Symptom
Multiple ring artifacts on axial scans - document in RFS - critical issue Multiple ring artifacts on axial scans
Error message
Error: 189057 Error in Vector Check Error in SD/Mean Error in Ratio !! See the LD for the full error code Error: 189057 Error in Vector Check Error in SD/Mean Error in Ratio !! Error: 189057 File: warmup_calc.cxx Error in Vector Check !! Error: 189057 File: warmup_calc.cxx Error in SD/Mean !! Error: 189057File: warmup_calc.cxx Line: 452 Error in Ratio !! Error: 189057 File: warmup_calc.cxx Line: 464 Error in HPF !! ! Error: 189060 Warmup ratio is out of range: 574ch. See Long Description for complete error Thu Jun 2 01:00:01 2005 189060 2 CTO1 OC0 ToolCtrl warmup errchk.c 177 Warmup ratio is out of range: 574ch.
Applies to
- Product
- CT/e, CT Prospeed AI/FI Irix or Linux See long description for sale information Product Name: CT/e Single CT Scanner: Irix Platform A list of types. (a Product/hardware table) N/A An On-site / Remote identification aid . (Example: check_config) N/A A Table of options (included, optional, new introductions, …) N/A A list of Software that has run on this product (Product / Software Compatibility Table) 3.01 / 3.02 / 3.03 Linux CTe Product photographs N/A Marketing Information N/A Documents: Always use the most appropriate documentation for your system. Documentation can be found on CT Service CDROM/DVD, or online at FCT Service Engineering Tech Pubs Website, Common Document Library, and My Workshop. Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found in the attachment Always use the most appropriate documentation for your system.
- Component
- 2138808 bowtie filter 2138808 bowtie filter Part #: 2138808 Part Type: GP CONSUMABLES Part Weight: 3.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: P9234MW FILTER 2138808 bowtie filter Part #: 2138808 Part Type: GP CONSUMABLES Part Weight: 3.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: P9234MW FILTER Comments: FLS - BOW TIE FILTER FOR YMS HISPEED& ESR CT MS7368 01/10/00
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.