PDU Main Breaker Trips, Will Not Reset due to Failed TVSS Module.

Symptom

Phoenix PDU Main Breaker Trips, Will Not Reset

Applies to

Models
Ovation
Product
Ovation/MFO
Software
MFO2 Ovation2 with Octane2
Component
2185383-11 MOV MODULE Also known as "TVSS". When FE's attempt to order the TVSS module listed in HFO3 Renewal Parts: 2228875-7, FEMC automatically "changes up" to p/n 2185383-11 MOV MODULE instead, because 2228875-7 cannot be obtained anymore. Phoenix is out of business. I checked with Margie Wiza of MR Eng and this is as intended. This is to fix PQR # 13109044.