I-Cross Over-Current Fault Due to XGD Amplifier.

Symptom

Will Not Scan with Gradient Errors.

Error message

2267868 The % axis XGA is reporting an I-Cross Over-current Fault See Long Description Error: 2267868 Host: MRS1_OC0 Process: NSP File: SCP:FspMac.cpp Line: 320 Exception Class: Unknown Severity: Unknown Function: Gradient: N/A PSD: Spin Echo Coil: 8HRBRAIN Scan: 926/1 The X axis XGA is reporting an I-Cross Over-current Fault. Extended Error : Ensure that the corresponding XPS is not reporting any output voltage faults on the LV bus. Retry the scan and if the fault persists, replace the XGA.

Applies to

Models
Discovery
Product
Discovery MR750. Also called DVMR. This product is compatible with 3.0T 20.0, 20.1_IB1, IB2, 21.0 and DV22.0 software only
Software
20.1_IB1 and IB2 This release of software is applicable to Discovery MR450 and Discovery MR750 systems. See Long Description 20.1_IB1 is the unified Discovery MR450 and Discovery MR750 software release. Known internally to engineering as "DVIB". This release of software is applicable to Discovery MR450 and Discovery MR750 systems. 20.1_IB2 is the upgrade (bug fixes) for IB1.
Component
5184566-2 XGD AMPLIFIER 5184566-2 XGD AMPLIFIER
Area
MR

Full cause & step-by-step resolution