GE Healthcare · MRDiscovery2267889
Scan pause and gradient error 2267889 due to failed XGD Amplifier.Scan Pauses with Gradient Errors and Continues after s…
Scan pause and gradient error 2267889 due to failed XGD Amplifier.Scan Pauses with Gradient Errors and Continues after start Scan Pressed then Images Show Bright Line Artifact
Symptom
Scan Pauses with Gradient Errors and Continues after start Scan Pressed then Images Show Bright Line Artifact System scans and pauses with Gradient errors. Images had a definite gradient collapsed.
Error message
2267889 The Z axis XPS is reporting a High Voltage I-Bus Over-current Fault See Long Description Thu Apr 16 08:42:00 2009 Error: 2267889 Host: MR3T Process: NSP File: SCP:FspMac.cpp Line: 320 Exception Class: Unknown Severity: Unknown Function: Gradient: N/A PSD: sprlio Coil: HEAD Scan: 193/7 The Z axis XPS is reporting a High Voltage I-Bus Over-current Fault.-> Extended Error: Ensure that the XPS and XGA are properly connected, and that the XGA is operating properly by swapping XPS modules if necessary. If fault persists replace the failing XPS
Applies to
- Models
- Discovery
- Product
- Discovery MR450 Also called DVMR. This product is compatible with 1.5T 20.1, 20.1_IB1, IB2, 21.0 and DV22.0 software systems only.
- Software
- 20.1 Also called DVMR. This software is compatible with 1.5T Discovery MR450 systems only. This version is not to be confused with the "20.1_IB1" or "20.1_IB2" releases.
- Component
- 5184566-2 XGD AMPLIFIER 5184566-2 XGD AMPLIFIER
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA ID100513. Will Not Scan with Gradient Errors.
- Corrupt Image Database.
- CWA ID103384. Computer Will Not Boot with the message on the screen that says Do Not Interact With System.
- CWA ID103582. Unable To Scan With Driver Module Errors.
- CWA ID104077. Computer Will Not Boot with the message on the screen that says Do Not Interact With System.
- CWA ID104279. System Software suddenly locked up.