Scan pause and gradient error 2267889 due to failed XGD Amplifier.Scan Pauses with Gradient Errors and Continues after s…

Scan pause and gradient error 2267889 due to failed XGD Amplifier.Scan Pauses with Gradient Errors and Continues after start Scan Pressed then Images Show Bright Line Artifact

Symptom

Scan Pauses with Gradient Errors and Continues after start Scan Pressed then Images Show Bright Line Artifact System scans and pauses with Gradient errors. Images had a definite gradient collapsed.

Error message

2267889 The Z axis XPS is reporting a High Voltage I-Bus Over-current Fault See Long Description Thu Apr 16 08:42:00 2009 Error: 2267889 Host: MR3T Process: NSP File: SCP:FspMac.cpp Line: 320 Exception Class: Unknown Severity: Unknown Function: Gradient: N/A PSD: sprlio Coil: HEAD Scan: 193/7 The Z axis XPS is reporting a High Voltage I-Bus Over-current Fault.-> Extended Error: Ensure that the XPS and XGA are properly connected, and that the XGA is operating properly by swapping XPS modules if necessary. If fault persists replace the failing XPS

Applies to

Models
Discovery
Product
Discovery MR450 Also called DVMR. This product is compatible with 1.5T 20.1, 20.1_IB1, IB2, 21.0 and DV22.0 software systems only.
Software
20.1 Also called DVMR. This software is compatible with 1.5T Discovery MR450 systems only. This version is not to be confused with the "20.1_IB1" or "20.1_IB2" releases.
Component
5184566-2 XGD AMPLIFIER 5184566-2 XGD AMPLIFIER
Area
MR

Full cause & step-by-step resolution