GE Healthcare · CT186427186436186421
CT/e, CT Prospeed AI/FI Irix or Linux
Symptom
Scan paused WITH Jedi / OGP communication errors. Scan paused WITH Jedi fails to talk to OGP Error: 186436 OGP Event Error.Code : 20-1030-01 Error detected by : OGP OGP did not receive any reply from JEDI in response to Rotor ON/OFF request
Error message
Error: 186427 Code : 20-0014-06 Some retries happened on CAN bus between JEDI and OGP and CAN bus becomes WARNING Status as the result. See LD for complete error Fri Sep 30 08:49:59 2005 Error: 186427 Host: CTABCCC Process: tgp_out File: tgp_errors.c Line: 896 XG Proc Comm Error. Code : 20-0014-06 Infomation detected by : OGP Some retries happened on CAN bus between JEDI and OGP and CAN bus becomes WARNING Status as the result. Error: 186436 Code : 20-1030-01 OGP did not receive any reply from JEDI in response to Rotor ON/OFF request Error: 186436 Host: GIMCCT Process: tgp_out File: tgp_errors.c Line: 395 OGP Event Error. Code : 20-1030-01 Error detected by : OGP OGP did not receive any reply from JEDI in response to Rotor ON/OFF request Error 186421 : Safety Loop Open 11-10F2-01 See long description for full error Thu May 29 17:33:45 2008 Error: 186421 Host: philact1 Process: tgp_out File: tgp_in_input.c Line: 1241 Safety Loop Open Code : 11-10F2-01 Error detected by : TGP(MP) mA/kV status was not received from OGP during X-ray ON. Or X-ray OFF Status was not received correctly from OGP. Safety Loop was opened by TGP as the result.
Applies to
- Product
- CT/e, CT Prospeed AI/FI Irix or Linux See long description for sale information Product Name: CT/e Single CT Scanner: Irix Platform A list of types. (a Product/hardware table) N/A An On-site / Remote identification aid . (Example: check_config) N/A A Table of options (included, optional, new introductions, …) N/A A list of Software that has run on this product (Product / Software Compatibility Table) 3.01 / 3.02 / 3.03 Linux CTe Product photographs N/A Marketing Information N/A Documents: Always use the most appropriate documentation for your system. Documentation can be found on CT Service CDROM/DVD, or online at FCT Service Engineering Tech Pubs Website, Common Document Library, and My Workshop. Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found in the attachment Always use the most appropriate documentation for your system.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required