Pular para o conteúdo principal
GE Healthcare · MRDiscovery226882220371472037168

After land marking and pushing “Advance to Scan” button the cradle travel position is lost on the enclosure display and…

After land marking and pushing “Advance to Scan” button the cradle travel position is lost on the enclosure display and error 2268822 is seen in the error log due to residue accumulation around brass…

Symptom

After land marking and pushing “Advance to Scan” button the cradle travel position is lost on the enclosure display and error 2268822 is seen in the error log. See Long Description 2268822 ERROR: EM_SRI_CAN_MOVE_DIDN'T_START_HOME Cradle didn't start moving when requested. Home switch may not be functioning. Last stop position: -2037147 um Current position: -2037168 um-> Extended Error: The move didn't start when requested. This may be due to several factors. The cradle may be obstructed. Check the error log for errors indicating a cable is missing. If the motor encoder was changing or if the motor controller reported a fault a different error would have been logged. If the cradle was already in the home position, this error may have also been caused by a faulty home switch in the LPCA.

Error message

2268822 Cradle didn't start moving when requested. Home switch may not be functioning. See Long Description 2268822 ERROR: EM_SRI_CAN_MOVE_DIDNT_START_HOME Cradle didn't start moving when requested. Home switch may not be functioning. Last stop position: -2037147 um Current position: -2037168 um-> Extended Error: The move didn't start when requested. This may be due to severalfactors. The cradle may be obstructed. Check the error log for errorsindicating a cable is missing. If the motor encoder was changing or ifthe motor controller reported a fault a different error would have beenlogged. If the cradle was already in the home position, this error mayhave also been caused by a faulty home switch in the LPCA.

Applies to

Models
Discovery
Product
Discovery MR750. Also called DVMR. This product is compatible with 3.0T 20.0, 20.1_IB1, IB2, 21.0 and DV22.0 software only
Software
20.0 Also called DVMR. This software is compatible with 3.0T Signa MR750 systems only
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues