Will Not Scan with Gradient Errors.

Symptom

Will Not Scan with Gradient Errors.

Error message

2267885 The % axis XPS is reporting a PLL Warning See Long Description Error: 2267885 Host: signa2 Process: NSP File: SCP:FspMac.cpp Line: 320 Exception Class: Unknown Severity: Unknown Function: Gradient: N/A PSD: Unknown Coil: Unknown Scan: -1/-1 The % axis XPS is reporting a PLL Warning.-> Extended Error: Ensure the fiber optic cable J4 is connected to the corresponding XPS, and that light is present. Check that the corresponding channel on the XPS-CB in the Cronus chassis has a good fiber optic output.

Applies to

Models
Discovery
Product
Discovery MR750. Also called DVMR. This product is compatible with 3.0T 20.0, 20.1_IB1, IB2, 21.0 and DV22.0 software only
Software
20.0 Also called DVMR. This software is compatible with 3.0T Signa MR750 systems only
Area
MR

Full cause & step-by-step resolution