GE Healthcare · MRDiscovery2268818507822268819
Display for cradle position goes from 2381 to '"No Home", even though the cradle is ~ 12 inches from end of travel. due…
Display for cradle position goes from 2381 to '"No Home", even though the cradle is ~ 12 inches from end of travel. due to Scan Room size selection in Guided Install.
Symptom
When you drive the table in, the display for cradle position goes from 2381 to No Home, even though the cradle is ~ 12 inches from end of travel You then need to pull it all the way to home to reset.
Error message
2268818 Reached position past expected End of travel switch however no EOT switch received Mon Jan 18 08:45:10 2010 Error: 2268818 Host: LIBR-MR1 Process: NSP File: SCP:LongMotionMonit Line: 447Exception Class: Unknown Severity: UnknownFunction: Patient Handling: Control/Display PSD: fsespt Coil: BODY Scan: 50782/1Reached position past expected End of travel switch however no EOT switch received. Press the home button to re-set home position. position : 2532399um-> No Home See Long Description " No Home " This message is displayed on the In Room Monitor (IRM) screen. It isn't logged anywhere. 2268819 Velocity tolerance exceeded Mon Jan 18 08:45:10 2010 Error: 2268819 Host: LIBR-MR1 Process: NSP File: SCP:LongMotionMonit Line: 1995 Exception Class: Unknown Severity: Unknown Function: Patient Handling: Control/DisplayPSD: fsespt Coil: BODY Scan: 50782/1Velocity tolerance exceeded.Expected: 100000um/sActual: 16600um/s
Applies to
- Models
- Discovery
- Product
- Discovery MR750. Also called DVMR. This product is compatible with 3.0T 20.0, 20.1_IB1, IB2, 21.0 and DV22.0 software only
- Software
- 20.0 Also called DVMR. This software is compatible with 3.0T Signa MR750 systems only
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA ID100513. Will Not Scan with Gradient Errors.
- Corrupt Image Database.
- CWA ID103384. Computer Will Not Boot with the message on the screen that says Do Not Interact With System.
- CWA ID103582. Unable To Scan With Driver Module Errors.
- CWA ID104077. Computer Will Not Boot with the message on the screen that says Do Not Interact With System.
- CWA ID104279. System Software suddenly locked up.