IRF3 is Not Ready Due to PDU 48V power supply.

Symptom

Cannot reset TPS with IRF3 not ready message Cannot reset TPS with IRF3 not ready message.

Error message

2267431 The IRF3 is not ready See Long Description Error: 2267431 Host: MR02 Process: NSPFile: AGP:Irf3RrfHWO Line: 1209Exception Class: Unknown Severity: UnknownFunction: Software/Firmware/Config: N/APSD: Unknown Coil: Unknown Scan: -1/-1 The IRF3 is not ready.-> Extended Error: This is just a warning message that indicates that another error message has already been logged about a serious problem with the IRF3, DTX, RRX, TDM, interconnect cable, or other error that is preventing the system from controlling the exciters and receivers. Look for a previous error message about configuration errors, cable errors, module errors, or calibration errors. Scans will not be allowed until that problem is corrected

2267414 NO XGD MODULE FOUND. The IRF3 board does not see the XGD module. See Long Description Error: 2267414 Host: PHSRADMR02 Process: diagCoordFile: Irf3.cpp Line: 2233Task ID: 0x493f570 Time: 25411 Object: AGPException Class: NotificationFunction: Software/Firmware/Config: N/A NO XGD MODULE FOUND.The IRF3 board does not see the XGD module. The system must have a dual fiber cable between the GMC board (in the Cronus Chassis) of the XGD and the IRF3 board-> Extended Error : The IR

Applies to

Models
Discovery
Product
Discovery MR750. Also called DVMR. This product is compatible with 3.0T 20.0, 20.1_IB1, IB2, 21.0 and DV22.0 software only
Software
20.0 Also called DVMR. This software is compatible with 3.0T Signa MR750 systems only
Component
5140621-4 48V Power Supply 1000W Condor PN GPMP 900_48 TEAL PN 0870109 5140621-4 48V Power Supply 1000W Condor PN GPMP 900_48 TEAL PN 0870109. Part is being set up for 3.0T DVMR program RSPL. 501139061, 5-Mar-08.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required