Pular para o conteúdo principal
GE Healthcare · CTDiscovery26010015217939

Scanning System Aborted Scan

Symptom

Scanning System Aborted Scan. Scanning System Aborted Scan

Error message

Error: 260100152 Exposure Interlock Opened Please refer to the long description for further details. Fri Feb 8 18:12:19 2008 Error: 260100152 Host: tgp Process: tgp_subsys File: scIsrService.c Line: 386 Fri Feb 8 18:12:19 2008 Host : Table Gantry Processor Ermes # : 260100152 Exception Class : Abort Severity : Primary File : scIsrService.c Line# : 386 Function : System Monitoring : Xray Exposure Interlocks Scan Type : Helical Scan: 17939/3/1 Exposure Interlock Opened

Applies to

Models
Discovery
Product
Discovery CT750 HD V40_PS_HD64_G_GTL V40_PS_HD64_G_GTM Discovery HD CT750 Method of visual identification HD gantry: Gray top and Gray Side covers with HD Markings Console Type: GOC6 has black SCIM and tower face Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of softwares the product can run (Consult attached FCT Software Cross Reference): 08MW39.4 - M3 (see attachement for M3 available options) 09MW29.14 - M4 (HD Options = M3 plus Gemstone Spectral Imaging, Gemstone Spectral Review and volume Helical Shuttle ) Example: Of difference with options software [insite@ct01 ~]$ swokinstall -p Product Software Option Keys: Gemstone Spectral Imaging Permanent HelicalShuttle Permanent Xtream Recon Permanent FAQ's Table weight limit: 500 lbs Tube Type - D3702T Green D3701T Current production - M4 Refer to product data sheets or Technical reference Manual, English 5317222-1EN Revision : 2 or later Refer to Application Tips and Work-Arounds Direction 5311815-1EN Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
5855030 (replaces 5184738) HD DIP PWA - (aka DIG DIP board) - used on a CT750-HD system. See Long Description for part details HD DIP board Part #: 5855030 This board is used on a CT750-HD system. Part #: 5855030 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: DFM type HDDIP PWA FRU: Yes Status: Orderable Old FRU Compatible Yes Comments added per to Doc0770989 by SSO# 305024364 in Nov. 4th Additional FRU Needed No Old FRU Scrapped No Old FRU Upgradable No Use on Hands Yes Use Demand History Yes Ultimate Prime 5855030 New FRU Backward Compatible Yes Old FRU Usable Yes Change Type Indirect ECO Number Doc0770989 Part #: 5184738 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: HD DIP PWA FRU: Yes Status: Orderable
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.