GE Healthcare · MRDiscovery2269059
Missing 80MHZ clock signal due to failed NB exciter board.
Symptom
Fails TPS Reset when booting up to Applications or when Resetting TPS See Long Description When the MR system is booting up to applications it will fail TPS reset but Applications will successfully boot. With Applications up and "Reset TPS" has been selected, TPS fails to download successfully.
Error message
2269059 The 1-Wire Network did not detect the Clock Reference 80MHz Clock that is generated in the Exciter Module 2269059 Task ID: 0x6db4420 Time: 4251 Object: AGPException Class: Notification Function: Software/Firmware/Config: N/A The 1-Wire Network did not detect the Clock Reference 80MHz Clock that is generated in the Exciter Module.->
Applies to
- Models
- Discovery
- Product
- Discovery MR450 Also called DVMR. This product is compatible with 1.5T 20.1, 20.1_IB1, IB2, 21.0 and DV22.0 software systems only.
- Software
- 20.1 Also called DVMR. This software is compatible with 1.5T Discovery MR450 systems only. This version is not to be confused with the "20.1_IB1" or "20.1_IB2" releases.
- Component
- 5154662 1.5T Reference Clock and HDv NB exciter Module 5154662 1.5T Reference Clock and HDv NB exciter Module.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA ID100513. Will Not Scan with Gradient Errors.
- Corrupt Image Database.
- CWA ID103384. Computer Will Not Boot with the message on the screen that says Do Not Interact With System.
- CWA ID103582. Unable To Scan With Driver Module Errors.
- CWA ID104077. Computer Will Not Boot with the message on the screen that says Do Not Interact With System.
- CWA ID104279. System Software suddenly locked up.