GE Healthcare · CT2300065292508823000904123000301223000652830771
Lightspeed 1.X - Octane
Symptom
Recon Status Indicates: Recon Resetting; Intermittent recon retries; Image recon delay noticed by cust See Long Description. Not for GRE Recon Status Indicates: Recon Resetting; Intermittent recon retries; Image recon delay could sometimes be noticed by customer. Issue could also cause SmartPrep exams to seem to hang. IG DSP encountered scan data view checksum error and the iceConsole.log posted repeated RECON_IG_RESET messages. Issue and errors can be very intermittent.
Error message
Error: 230006529 Recon Mgr is sending queue ack to queuemgr with a RECON_JOB_FAILED_RETRY status See Long Description for Full Error Tue Feb 1 06:33:34 2005 Error: 230006529 Host: ice Process: reconmgr File: rm_recon_events.c Line: 329 Function: Image Generation : SBC Processing Job Type: Prospective Axial Scan: 25088/2/9 Image: 25088/3/25 Recon Seq Id: 957 Recon Mgr is sending queue ack to queuemgr with a RECON_JOB_FAILED_RETRY status Error: 230009041 The image did not complete reconstruction within the time allotted. See long description for complete error Tue Apr 15 19:45:47 2003 Error: 230009041 Host: smopct1 Process: queuemgr File: qmgr_queue_cntl.c Line: 2524 Function: Image Generation : SBC Processing Job Type: Retro Axial Scan: 414/4/1 Image: 414/4/5 The image did not complete reconstruction within the time allotted. Error: 230003012 IG unit 0 failed to reset successfully. See long description for full error Tue Sep 9 08:19:33 2008 Error: 230003012 Host: ct2l sbc Process: igctrl File: igctrl mgr. Line: 437 IG unit 0 failed to reset successfully. Retried 2 times. Possible hardware problem. No further attempts to use this IG will be made until the system is shutdown and res
Applies to
- Product
- Lightspeed 1.X - Octane 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete Lightspeed 1.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config showprods|grep Apps note: if check_config does not work then the software is not up to date with all FMI's. Check FMI history and verify that all patches have been loaded since last software reload. Ref FMI 25269 List of Supported Console Types Octane with Pegasus Recon hardware. may have an 02 SBC - Run df -k to see if the sbc is partitioned. List of softwares the product can run 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3176T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2118899-2 RHAPSODE IMAGE GENERATOR Board (IG Board). Used in Lightspeed 1x, CTi 4x, CTi 6x systems. See LD for details Part #: 2118899-2 Part Type: GP REPAIRABLES Part Weight: 3.64 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: RHAPSODE IMAGE GENERATOR FRU: Yes Status: Orderable Comments: FLS - IMAGE GENERATOR BOARD USED ON RHAPSODE MS1356 BURR 12-95 REPLACED 2118899 PER PCN 189598. MS4636 2/6/96 PCN 189939 GELS - KANBAN CSS F00,PCB IMAGE GENERATOR
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.