Pular para o conteúdo principal
GE Healthcare · MROvation22402651177491864

Will Not Scan with Gradient Errors.

Symptom

Will Not Scan with Gradient Errors.

Error message

2240265 Y Axis Gradient Amplifier Fault. 2240265 1 NSP SCP:GradientScaMdsHWO 1201 Y Axis Gradient Amplifier Fault. RMS current overload. Possible causes: PSD is exceeding the limits of the Gradient Subsystem or faulty High Speed Serial Data (fiberoptic) cables. Please Reset TPS to recover the Gradient Subsystem. EN 339 SR 335 1177491864 1 1

Applies to

Models
Ovation
Product
Ovation with EXCITE
Software
OPENMR4-MFO4 Applies to Ovation4 (Excite) only. Ovation4 (MFO4) is 11.x based software.
Component
5118227 AN8295Y Analogic Gradient Amplifier Unit 5118227 AN8295Y Analogic Gradient Amplifier Unit 5118227 AN8295Y Analogic Gradient Amplifier Unit
Area
MR