Pular para o conteúdo principal
GE Healthcare · MRDiscovery221853722720352218616

Several Series Were Missing and Never Reconstructed.

Symptom

Several Series Were Missing and Never Reconstructed. Server DP Internal Queue Full and Scanner State Machine Errors Logged. See Long Description and Attachment Discovery MR750 system scanned exam 1070 and a server error occurred requesting TPS reset. Missing out of 10 series were series numbers 5, 6, 8, 9, 10. Series 1, 2, 3, 4 and 7 reconstructed. Server queue full and scanner state machine errors were found logged when this incident occurred. See Attachment for the errors that were logged. iTrak # 13274795.

Error message

2218537 Scan stopped, Scan Stopped due to error Internal data queue full. See Long Description 2218537 ../MsaScanMgr.m 3168 scn Scan stopped, Scan Stopped due to error : Internal data queue full. Most likely the application is not keeping up with the incoming acquisition data rate..

2272035 Server Name: DP Internal Control data queue full See Long Description 2272035 AcqHWO.cpp 534 LxMxConsumer Server Name: DP Internal Control data queue full. Most likely the application on DP is not keeping up with the incoming acquisition data rate.

2218616 Internal software error has occurred in scanner state machine. The scanner cannot accept start scan event, when in prescanning state. See Long Description 2218616 ./ScannerState.c scn Internal software error has occurred in scanner state machine. The scanner cannot accept start scan event, when in prescanning state.

2272035 Server Name DP Internal Frame data queue full. See Long Description 2272035 AcqHWO.cpp 534 LxMxConsumer Server Name: DP Internal Frame data queue full. Most likely the application on DP is not keeping up with the incoming acquisition data rate.

Applies to

Models
Discovery
Product
Discovery MR750. Also called DVMR. This product is compatible with 3.0T 20.0, 20.1_IB1, IB2, 21.0 and DV22.0 software only
Software
20.0 Also called DVMR. This software is compatible with 3.0T Signa MR750 systems only
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.