Part Numbers for the Filters used on the Rear of the SSM.
Symptom
Part Numbers for the SSM Filters is Difficult to Find See Long Description There are 4 filters installed on the back of the SSM; 2-25 pin, 1-37 pin and 1-15 pin DB Filters. Missing filters can cause a drop in SNR due to higher noise values. This is due to a normal amount of noise from the power supplies in the SSM. Missing filters happens when the SSM is replaced and the filters are not transfered from the old unit to the new unit. Replacement SSM's do not include the filters.
Applies to
- Models
- Ovation
- Product
- Ovation/MFO
- Software
- MFO3 Applies to Ovation3 with HP4000 computer running Linux OS.
- Area
- MR
Full cause & step-by-step resolution