2224667 Manual Prescan failed, RF Hub RF/bias setup not complete due to defective RF Hub Control Board.

Symptom

Unable to Scan with RF Hub RF/Bias Setup not Complete 2224667 1 bia5 psc../PscMainControl.cc 1409Exception Class: UnknownFunction: Prescan: Host Manual PrescanManual Prescan failed, RF Hub RF/bias setup not complete

Error message

2274008 Multicoil bias failure RF Hub delayed unblank pulse delay out of range. 2274008 2 bia5 NSPSCP:Bias 2880Exception Class: Unknown Severity: UnknownFunction: RF: RF BiasPSD: epi_service Coil: BODY Scan: 50001/1Multicoil bias failure.RF Hub delayed unblank pulse delay out of range.Board: RF Hub Slot 11 (RFCB)Count: 1Delay: * 21600 ns *Width: 2700 ns->

2224667 Manual Prescan failed, RF Hub RF/bias setup not complete. 2224667 1 bia5 psc../PscMainControl.cc 1409Exception Class: UnknownFunction: Prescan: Host Manual PrescanManual Prescan failed, RF Hub RF/bias setup not complete

Applies to

Models
Discovery
Product
Discovery MR750. Also called DVMR. This product is compatible with 3.0T 20.0, 20.1_IB1, IB2, 21.0 and DV22.0 software only
Software
20.0 Also called DVMR. This software is compatible with 3.0T Signa MR750 systems only
Component
5250098-2 3.0T RF HUB CONTROL BOARD 5250098-2 3.0T RF HUB CONTROL BOARD. Comments: MUST run all receiver diagnostics, including Coil Datapath Diagnostic, prior to placing order for this part. 212002950 October 24, 2008. Part is being set up for 3.0T DVMR program RSPL. 501139061, 5-Mar-08.
Area
MR

Full cause & step-by-step resolution