Troubleshooting Steps for Scan Stops Intermittently with Error Code 2254614 RF Amplifier High Voltage High Fault.

Symptom

Scan Stops Intermittently with Error Code 2254614 RF Amplifier High Voltage High Fault. RF Amplifier high voltage power supply voltage went above specified maximum voltage. How do we troubleshoot this problem? See Long Description Scan stops intermittently with error code 2254614 "RF Amplifier High Voltage High Fault. RF Amplifier high voltage power supply voltage went above specified maximum voltage." How do we troubleshoot this problem? 2254614 ERROR: EM_RF_XRFD_ERROR_46 RF Amplifier High Voltage High Fault. RF Amplifier high voltage power supply voltage went above specified maximum voltage. HV: %.2f V, Upper spec: %.2f V, Unit:%d(1,2->PS1, PS2)

Error message

2254614 RF Amplifier High Voltage High Fault. RF Amplifier high voltage power supply voltage went above specified maximum voltage. See Long Description 2254614 ERROR: EM_RF_XRFD_ERROR_46 RF Amplifier High Voltage High Fault. RF Amplifier high voltage power supply voltage went above specified maximum voltage. HV: %.2f V, Upper spec: %.2f V, Unit:%d(1,2->PS1, PS2)

Applies to

Models
Discovery
Product
Discovery MR750. Also called DVMR. This product is compatible with 3.0T 20.0, 20.1_IB1, IB2, 21.0 and DV22.0 software only
Software
20.0 Also called DVMR. This software is compatible with 3.0T Signa MR750 systems only
Area
MR