System Stops Scanning due to Error 2275652 "RF Hub detected a fault while scanning" due to Fiber Optic Cable Run 2025-D…

System Stops Scanning due to Error 2275652 "RF Hub detected a fault while scanning" due to Fiber Optic Cable Run 2025-D Coiled, Bent, or Kinked into a Diameter Less Than 12 Inches (30.5 cm.).

Symptom

System Stops Scanning with Error 2275652 RF Hub detected a fault while scanning. See Long Description System scanning suddenly stops and only message reported in error log at time scan stopped is reported below: 2275652 ERROR: EM_RF_HUB_WAKEUP_STOP_SCAN RF Hub detected a fault while scanning.

Error message

2275652 RF Hub detected a fault while scanning See Long Description 2275652ERROR: EM_RF_HUB_WAKEUP_STOP_SCAN RF Hub detected a fault while scanning.

Applies to

Models
Discovery
Product
Discovery MR750. Also called DVMR. This product is compatible with 3.0T 20.0, 20.1_IB1, IB2, 21.0 and DV22.0 software only
Software
20.0 Also called DVMR. This software is compatible with 3.0T Signa MR750 systems only
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.