Pular para o conteúdo principal
GE Healthcare · MROvation

The PDU is in Stop Mode Due to Phoenix PDU control board.

Symptom

The PDU is in Stop Mode and the EMO Reset Will Not Work See Long Description Sometimes when you push the green EMO reset button on the PDU nothing happens. You cannot hear the 24VDC relay click on nor hear any fans start up. In addition, you keep getting an error message that the PDU is in stop mode or the 24VDC supply has failed after you reset the TPS and attempt to scan.

Applies to

Models
Ovation
Product
Ovation/MFO
Software
MFO3 Applies to Ovation3 with HP4000 computer running Linux OS.
Component
2228875-2 Phoenix PDU Control Board (new part #2330230) 2330230 REPLACES 2228875-2, PARTS ARE FORWARD BACKWARD COMPATIBLE. 2228875-2 EOPL, replaced by 2330230 which is backward compatable
Area
MR

Full cause & step-by-step resolution